J Appl Biomed 11:27-32, 2013 | DOI: 10.2478/v10136-012-0016-9
Automated detection of organophosphate warfare gases (nerve agents) in air based on micro-SIA - lab-on-valve system
- 1 Faculty of Pharmacy, Charles University, Hradec Králové, Czech Republic
- 2 Faculty of Military Health Sciences, University of Defense, POB 35-T, Hradec Králové, Czech Republic
Equipment for fast and accurate detection of organophosphate nerve agents is developed and tested. The method is based on the spectrophotometric monitoring of the enzyme activity of butyrylcholinesterase after its contact with air in a special absorption unit (a "scrubber") developed for the purpose. The scrubber was made from a glass tube filled with glass beads (diam. 3 mm) and filled with approx. 5 ml of butyrylcholinesterase in a phosphate buffer of pH 7.4. The air sample was bubbled through this solution for 20 s at a flow rate of 80 l hour-1. Thereafter 8 μl of the enzyme solution were aspirated into the micro-SIA-LOV analyzer and the activity of the enzymes were evaluated by using Ellman's reagent, i.e. 2.5 mmol l-1 butyrylthiocholine iodide and 0.25 mmol 5,5'-dithiobis (2-nitrobenzoic acid). The absorbance of the coloured reaction product was measured at 412 nm after the reaction time of 60 s. The residue of the absorption liquid was washed away from the absorber and the system was washed with the enzyme solution prior to next analysis. The contaminated air caused partial inhibition of the enzyme activity of the absorption liquid. The activity of the contaminated sample was compared with the activity of the unaffected enzyme (blank measurement). The analysis was controlled by two PCs. The effect of the concentration of analyte in the absorption liquid on the enzyme activity was tested for 10-5-10-9 mol l-1 sarin. A single analysis (including the absorption step) took
Keywords: sequential injection analysis; lab-on-valve; organophosphate; cholinesterase; nerve agent; sarin
Received: February 24, 2012; Revised: April 2, 2012; Published: March 31, 2013 Show citation
ACS | AIP | APA | ASA | Harvard | Chicago | Chicago Notes | IEEE | ISO690 | MLA | NLM | Turabian | Vancouver |
References
- Ellman GL, Courtney DK, Andres V. A new and rapid colorimetric determination of acetylcholinesterase activity. Biochem Pharmacol. 7: 88-90, 1961.
Go to original source...
Go to PubMed...
- Maruko H, Sekiguchi H, Seto Y, Sato A. Detection performance of aspiration-type ion mobility spectrometer for chemical warfare agents. Bunseki Kagaku. 55: 191-197, 2006.
Go to original source...
- Matsushita K, Sekiguchi H, Seto Y. Performance of portable surface acoustic wave sensor array detector for chemical agents (in Japanese). Bunseki Kagaku. 54: 83-88, 2005.
Go to original source...
- Mesillakso M. Chemical Weapons Convention Chemicals Analysis. Sample Collection, Preparation and Analytical Methods. J. Wiley & Sons, Chichester, 2005, pp. 67-77.
Go to original source...
- OPCW. Convention on the Prohibition, Stockpiling and Use of Chemical Weapons and on their Destruction. Organisation for the Prohibition of Chemical Weapons, The Hague, 1994.
- Riegle DW, D'Amato AM. U. S. Chemical and Biological Warfare-Related Dual Use Exports to Iraq and their Possible Impact on the Health Consequences of the Gulf War. United States Senate, 103rd Congress, 2nd Session, May 25, 1994, http://www.gulfweb.org/bigdoc/report/appgb.html,downloaded: 7th Feb 2012.
- Ruzicka J. Lab-on-valve: Universal microflow analyzer based on sequential and bead injection. Analyst. 125: 1053-1060, 2000.
Go to original source...
- Ševelová L, Vachek J, Bajgar J. Inhalation apparatus for generating sarin and soman toxic vapors. Acta Medica (Hradec Králové). 46: 157-161, 2003.
Go to original source...
Go to PubMed...